HPLC SYSTEM SUITABILITY VIRTUAL LABInstrument Control Room • Training Scenario SST-04
Pre-run qualificationSST FAILEDMethod-specific training criteria
QC Laboratory • HPLC Bay 02 • Instrument HPLC-07

System Suitability Dashboard

Critical-pair resolution1.62
Tailing factor1.28
Theoretical plates6,870
Area %RSD (6 std inj.)0.18%
Training method IMP-HPLC-042 Rev 7 defines these scenario-specific criteria: Resolution ≥ 2.00; Tailing ≤ 2.00; Plates ≥ 3,000; Standard area %RSD ≤ 2.00%. These are not universal HPLC limits.

Current gate

FAILED — critical-pair resolution does not meet the approved method criterion. Sample injections have not started.

Chromatogram Viewer

Impurity DAPI Retention timeResponse
Critical pairImpurity D / API
Calculated resolution1.62

Sequence Table

#InjectionTypeStatusKey note
1–6SST_STD_01–06Qualified SST standardCompleted / FAILEDCritical-pair Rs = 1.62
7BLANK_01BlankNot startedSequence held
8–13SAMPLE_01–06SamplesNot startedProtected by sequence hold

Approved Analytical Procedure

Method IDIMP-HPLC-042 Rev 7
ColumnC18, 150 × 4.6 mm
Flow1.000 mL/min
Detection254 nm
Column temperature35.0°C
SST materialQualified reference mixture

Method-specific SST requirements

ParameterCriterionCurrent
Critical-pair resolution≥ 2.001.62
Tailing factor≤ 2.001.28
Theoretical plates≥ 3,0006,870
Standard area %RSD≤ 2.00%0.18%

CDS Audit Trail

TimeUserEventContext
07:42analyst.rkSequence template copiedTemplate: IMP_HPLC_042_R6_MONTHLY
07:44analyst.rkAcquisition method loadedIMP_HPLC_042_R6 — column oven 30.0°C
08:13analyst.rkSST sequence started6 standard injections
08:52systemSequence gate failedCritical-pair resolution below method criterion

Instrument System Log

ModuleCurrent stateLast event
Pump1.000 mL/min • 126 barNo alarms
AutosamplerTray 1 • 8°CNo alarms
Column oven30.0°C stableSetpoint loaded from acquisition method
UV detector254 nmLamp energy within internal check